Examining the cause of CT artifacts, isolating ring artifact, streak artifact and steps you can take to prevent CT system down time.

What is a CT Ring Artifact?

Ring artifacts in CT scans are circular or ring-like image distortions characterized by black or white edges that can degrade the diagnostic quality of images and can lead to misinterpretation or misdiagnosis.

What Causes CT Ring Artifacts?

These image distortions are produced when one or more detector channels generate spikes or dips in the scan data. A ring artifact appears in images when image data in one or more detector channels registers outside of the system’s tolerance levels. The most common causes of ring artifacts in CT scans are faulty detector components, noisy power supplies, calibration errors, obstructions in the X-ray beam, or even physical damage to detector modules. Read more below.

Note: Perform axial test scans using water phantom for most consistent scan data. Helical scans may show partial ring or artifact that may increase troubleshooting difficulty.

Perform test scans using water phantom and examine the nature of the artifact. Artifact descriptors will help in isolating cause of CT artifacts.

1. Induced CT Artifact

Description: Typically represented as band artifacts or ring artifact with blurred edges. Contrast present on mylar window around the CT bore or beam obstruction during daily calibration can be cause for artifact.

How to remove contrast from CT mylar window

First, inspect mylar window for any small holes in the ring. Do not apply liquid solution if holes are visible in mylar window. Then, use warm water and rubbing alcohol solution on a clean cloth to wipe off any visible contrast along beam path. In this case, pour a small amount of solution on cloth and wipe with cloth to avoid excessive liquid. Avoid spilling any solution into CT gantry.

FInally, after contrast has been cleaned, perform daily calibration and test scans using quality assurance phantom.

Beam obstructions during daily calibrations cause image artifact. Additionally, CT daily QA scans may show elevated mean values. Water mean value only passes specification if value is within -3 to +3. Perform daily calibrations again after identifying and clearing beam obstruction.

CT Scanner Ring Artifact

Description: Clearly defined edges and usually localized to channel segmentation no more than 1mm in thickness. White or black in color. Ring artifacts occur as a result of data acquisition or detector issues. If experiencing ring artifacts during scanning try:

1) Choosing a different detector configuration ex. 4×5, 2×10, 8×2.5. Also try to choose different collimator setting by reducing detector coverage in an attempt to exclude bad channel during scan data acquisition.

2) If you cannot scan with a configuration that is free of ring artifacts or excludes bad data channel, you most likely have a Data Acquisition subsystem issue. Most causes of white ring artifact and dark ring artifact will likely be within the CT Gantry and will require a Field Engineer.

How To Resolve CT Scan Ring Artifact

To resolve ring artifacts, troubleshooting and corrective action is required. This may involve removing obstructions in the path of the X-ray beam or replacing faulty detector components. After performing corrective action, the system will need to be recalibrated In order to resolve the CT ring artifact.

It is important to note that performing calibrations before addressing the underlying issue can mask the appearance of the ring artifact but will complicate troubleshooting the root cause.

To effectively resolve a CT scan ring artifact, it is advisable to conduct a thorough examination of the patient database to identify when the artifact first appeared and track the frequency of occurrence.

Once this information is obtained, promptly notifying the CT scan service provider with these details would be the recommended course of action. By doing so, the service provider can be informed of the artifact’s history and take appropriate measures to resolve the ring artifact effectively.

CT Ring Artifact Troubleshooting Tips

If you are currently experiencing a CT scan ring artifact, here’s a few things to know:

Ring artifacts in CT scans may not appear on every image and often occur only with specific scan parameters selected.

When ring artifacts are first noticed, it’s useful to run test scans with DQA phantoms to assess your system’s limitations.

Perform multiple test series, adjusting different scan parameters. Start by switching the focal point: if you’re using a large focal point, switch to a small one, or vice versa. Next, adjust detector configuration settings, such as slice thickness and images per rotation. You may find that only certain protocols are affected, or that all of them are. However, this quick 5-minute evaluation is invaluable for assessing current scan limitations and determining which patients can still be scanned without concerns about ring artifacts.

To learn more about CT scan ring artifacts, check out some of the helpful resources listed below.

4 Items To Check Before Making a Service Call

CT Scan Room Temperature

First, check your scan room temperature. Ideal scan room temperature ranges from 68 degrees to 73 degrees Fahrenheit. Scanning should not be performed above 83 degrees F. Rising temperatures in scan room can overheat internal components and cause CT artifacts. Pause scanning until CT scan room temperature has normalized and is within recommended scanning temperature.

CT Gantry Exhaust Fan Functionality

Second, check CT gantry exhaust fans. Check the top of your CT gantry to make sure all fans are on. Heat buildup inside of the CT Gantry can cause damage to sensitive electrical components and power supplies. High gantry exhaust temperature relative to ambient temperature can be an indication that gantry cooling subsystem is not functioning properly.

Are fans not working? Check for radiant heat from gantry fan ports. Issues with power supplies or faulty fans can prevent heat from being evacuated from CT Gantry.

Power Down and Restart CT scanner

Contact MRI Repair Service Engineer (FE)

Call qualified Field Service Engineer for CT repair service. Contact us for a list of recommended service providers.

Cause of CT Artifacts: Conclusion

Most ring and band artifacts will require troubleshooting from a qualified field service engineer. While the cause of CT artifacts can vary, there are some simple steps that can be taken to ensure risk of induced artifact in calibration data is reduced to a minimum.

Additional items that will keep your system up and running: Changing HVAC or air conditioning filters regularly, regular planned maintenance, tracking daily quality control scans to ensure consistency and spot anomalies.

Additional Resources

Does Your System Require On-site CT Scanner Repair Service? Contact Us Today!

Read more on Larry’s author page.

Latest Articles

The information provided by MRIPETCTSOURCE (“we,” “us,” or “our”) on https://www.medicalimagingsource.com (the “Site”) is for general informational purposes only. All information on the Site is provided in good faith, however we make no representation or warranty of any kind, express or implied, regarding the accuracy, adequacy, validity, reliability, availability, or completeness of any information on the Site. UNDER NO CIRCUMSTANCE SHALL WE HAVE ANY LIABILITY TO YOU FOR ANY LOSS OR DAMAGE OF ANY KIND INCURRED AS A RESULT OF THE USE OF THE SITE OR RELIANCE ON ANY INFORMATION PROVIDED ON THE SITE. YOUR USE OF THE SITE AND YOUR RELIANCE ON ANY INFORMATION ON THE SITE IS SOLELY AT YOUR OWN RISK.

Amazon and the Amazon logo are trademarks of Amazon.com, Inc. or its affiliates.

Larry Lopez

Larry is a biomedical imaging specialist with more than 16 years of professional experience in MRI, CT, and PET system installation, calibration, quality assurance, and advanced troubleshooting. As the founder, digital creator, and lead author of MRIPETCTSOURCE, he produces educational content designed to elevate the skills of technologists, engineers, and imaging center operators. Larry also serves as the chief technical advisor and lead web developer for MedicalImagingSource.com, where he oversees the accuracy, technical depth, and clinical relevance of all published resources. His work integrates field expertise with clear, evidence-based explanations to support both professionals and patients. Connect with Larry on social media: LinkedIn | YouTube | X (Twitter) | Instagram | Pinterest | Facebook

XACT Imaging: The Best of X-ray and Ultrasound

A UCI research team pioneered X-ray-induced acoustic computed tomography (XACT), a novel imaging modality that…

1 week

Best MRI-Safe Jewelry & Piercing Retainers (What’s Actually Safe)

Home Learning Hub MRI MRI Patient Resources Best MRI-Safe Jewelry & Piercing Retainers (What’s Actually…

3 months

MRI Laser Positioning: Why MRI Scanners Use Class 2 Lasers for Patient Landmarking

Home Learning Hub MRI MRI Patient Resources MRI Laser Positioning: Why MRI Scanners Use Class…

3 months

CT Ring Artifacts Explained

Home Learning Hub CT CT Tech Resources CT Ring Artifacts Explained Everything CT technologists need…

3 months

How X-Rays Work

Home Learning Hub Patient Resources Radiology https://www.medicalimagingsource.com/how-x-rays-work X-rays have been an essential diagnostic tool for…

3 months

Medical Imaging Media Gallery | MRIPETCTSOURCE

Home Learning Hub Imaging Center Resources Licensable Media Medical Imaging Media Gallery | MRIPETCTSOURCE MRI,…

5 months